Por favor, use este identificador para citar o enlazar este ítem: http://dspace.espoch.edu.ec/handle/123456789/15021
Registro completo de metadatos
Campo DC Valor Lengua/Idioma
dc.contributor.advisorSáez Paguay, Miguel Ángel-
dc.contributor.authorArguello Albiño, Marlon Steven-
dc.date.accessioned2022-03-10T16:12:23Z-
dc.date.available2022-03-10T16:12:23Z-
dc.date.issued2021-03-02-
dc.identifier.citationArguello Albiño, Marlon Steven. (2021). Validación de los blindajes estructurales de las zonas controladas en el área de diagnóstico y producción de radiofármacos en la Unidad Técnica PET Ciclotrón del hospital de especialidades Carlos Andrade Marín. Escuela Superior Politécnica de Chimborazo. Riobamba.es_ES
dc.identifier.urihttp://dspace.espoch.edu.ec/handle/123456789/15021-
dc.descriptionEl objetivo de este estudio es validar los blindajes estructurales en las áreas tanto de diagnóstico como de producción de radiofármacos de la Unidad Técnica PET Ciclotrón, por lo cual se tomaron varios puntos de interés para monitorizarlos y verificar con el cálculo de blindaje y levantamiento radiométrico el blindaje estructural. Se utilizó protocolos internacionales en conjunto con equipos detectores de radiación ionizante, en este caso se usó un detector Geiger - Müller, marca Thermo Scientific RadEye modelo G-10 y una cámara de ionización marca Ludlum modelo 9DP. Las ecuaciones empleadas tanto para el cálculo del blindaje y levantamiento radiométrico se adaptan a las consideraciones propias del estudio; se lo hizo de forma manual y través de plantillas en Excel. Los resultados muestran que todas las barreras de las salas de la Unidad Técnica PET Ciclotrón poseen valores que están dentro del rango de tolerancia tanto en el cálculo de blindaje como en el levantamiento radiométrico, las salas con mayor tasa de dosis son el cuarto caliente y sala PET/CT, y las salas con menor tasa de dosis son los cuartos PEM, laboratorio de radiofarmacia y control físico – químico, sin embargo garantizan la seguridad y protección radiológica del POE, pacientes y público dentro de la Unidad. Se concluye que todos los blindajes estructurales son seguros y no permiten que se superen las restricciones de dosis. Se recomienda en este estudio respetar las técnicas de minimización de dosis (distancia, tiempo y blindaje) especialmente cuando se transporta el material radiactivo y cuando los pacientes están inyectados.es_ES
dc.description.abstractThe objective of this study is to validate the structural shielding in the diagnostic and radiopharmaceutical production areas of the PET Cyclotron Technical Unit, for which several points of interest were taken to monitor them and verify the structural shielding with the shielding calculation and radiometric survey. International protocols were used together with ionizing radiation detectors, in this case a Geiger - Müller detector, Thermo Scientific RadEye model G-10 and a Ludlum model 9DP ionization chamber were used. The equations used for both the shielding and radiometric survey calculations are adapted to the study's own considerations; this was done manually and through Excel templates. The results show that all the barriers of the rooms of the PET Cyclotron Technical Unit have values that are within the tolerance range both in the shielding calculation and in the radiometric survey, the rooms with the highest dose rate are the hot room and PET/CT room, and the rooms with the lowest dose rate are the PEM rooms, radiopharmacy laboratory and physical-chemical control, however, they guarantee the safety and radiological protection of the POE, patients and public within the Unit. It is concluded that all structural shielding is safe and does not allow dose restrictions to be exceeded. It is recommended in this study to respect dose minimization techniques (distance, time and shielding) especially when transporting radioactive material and when patients are injected.es_ES
dc.language.isospaes_ES
dc.publisherEscuela Superior Politécnica de Chimborazoes_ES
dc.relation.ispartofseriesUDCTFC;86T00121-
dc.rightsinfo:eu-repo/semantics/openAccesses_ES
dc.subjectCIENCIAS EXACTAS Y NATURALESes_ES
dc.subjectFÍSICAes_ES
dc.subjectPOSITRONESes_ES
dc.subjectCÁLCULO DE BLINDAJEes_ES
dc.subjectLEVANTAMIENTO RADIOMÉTRICOes_ES
dc.subjectMEDICINA NUCLEAR MOLECULARes_ES
dc.subjectPROTECCIÓN RADIOLÓGICAes_ES
dc.subjectRADIOFÁRMACOes_ES
dc.titleValidación de los blindajes estructurales de las zonas controladas en el área de diagnóstico y producción de radiofármacos en la Unidad Técnica PET Ciclotrón del hospital de especialidades Carlos Andrade Marínes_ES
dc.typeinfo:eu-repo/semantics/bachelorThesises_ES
dc.contributor.miembrotribunalYanchapanta Bastidas, Vilma Nohemí-
dc.rights.urihttps://creativecommons.org/licenses/by-nc-sa/3.0/ec/es_ES
Aparece en las colecciones: Biofísica